2 Displaying Data
44 Agilent 1000B Series Oscilloscopes User’s Guide
The following modes or functions are not available in X-Y format:
Automatic voltage or time measurements.
Cursor measurements.
Mask testing.
Math function waveforms.
Reference waveforms.
Zoomed time base display.
Displaying waveforms as vectors.
Horizontal position knob.
Trigger controls.
To view the sample rate
1 Press [Horiz].
2 In the Horizontal menu, the Sa Rate menu item displays the sample rate
used for the current horizontal scale setting.
See Also “Memory Depth and Sample Rate” on page 71.