3Capturing Data
66 Agilent 1000B Series Oscilloscopes User’s Guide
Overview of Sampling
To understand the oscilloscope’s sampling and acquisition modes, it is
helpful to understand sampling theory, aliasing, oscilloscope bandwidth
and sample rate, oscilloscope rise time, oscilloscope bandwidth required,
and how memory depth affects sample rate.
Sampling Theory
The Nyquist sampling theorem states that for a limited bandwidth
(band- limited) signal with maximum frequency f
MAX
, the equally spaced
sampling frequency f
S
must be greater than twice the maximum frequency
f
MAX
, in order to have the signal be uniquely reconstructed without
aliasing.
f
MAX
= f
S
/2 = Nyquist frequency (f
N
) = folding frequency
Aliasing
Aliasing occurs when signals are under-sampled (f
S
< 2f
MAX
). Aliasing is
the signal distortion caused by low frequencies falsely reconstructed from
an insufficient number of sample points.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies DSO1024A and is the answer not in the manual?

Agilent Technologies DSO1024A Specifications

General IconGeneral
BrandAgilent Technologies
ModelDSO1024A
CategoryTest Equipment
LanguageEnglish

Related product manuals