Step 19
If you change the electronics
,
the DC input cable
and
flat
keypad cable has to be
disconnected.
Step 20
DC cable (+) is at the right most position
The cable is marked with a (1) sign
The fan connector can be seen to the left most
position
Step 21
It is easier to engage the flat cable to the connector
receptacle by having a
0.5 mm spacer underneath.
Step 22
You can use a very small flat pair of pliers to
engage the cable.