Installation
3-2 Opus Operator's Manual
3.3.1 Space and Positioning
Space should be allocated with adequate ventilation and free airflow. The working
area for the system should be prepared according to the system dimensions
presented in 4.3.3. In order to guarantee proper ventilation, always keep the sides
of the system at least 0.5 meters from the wall or from other obstructions to air
flow.
The system can interfere with computers and other electrical and medical systems.
It should be placed at least two meters away from any such systems.
Caution
In order to guarantee safe disconnection of the system in any hazardous situation
system shall be allocated in such a way that appliance coupler always accessible.
3.3.2 Electrical Requirements
The power supply allows the system to be plugged into any wall outlet that
supplies: 120VAC, 50/60Hz, 6.3A
Input power lines should be free of transients, voltage and current spikes, sags and
surges. Consequently, the system power line should not be shared with other heavy
variable loads such as elevators, air conditioning systems, large motors, etc.
The system is grounded through the grounding conductor in the power cable that
is plugged into the wall power outlet. Good grounding is essential for safe
operation.
Only use the power cable supplied with the system.
Caution
Verify that the system is wired for the appropriate line voltage of your country.
3.3.3 Environmental Requirements
Air Quality:
The system should operate in a non-corrosive atmosphere. Corrosive materials such
as acids can damage electrical wiring, electronic components and the surfaces of
RF delivery components.
Temperature:
To ensure that the system performs optimally, maintain room temperature between
20°C and 25°C and a relative humidity of less than 80%.