Opus Operator's Manual 7-1
Maintenance
7.1 Introduction
This chapter contains maintenance instructions for the system.
Routine maintenance may be performed by the beauty operator unless otherwise
specified. Any maintenance procedure not mentioned in this chapter must be
performed only by Alma Lasers authorized technical personnel.
The system is designed to operate reliably without any need for operator
maintenance. However, the outer surfaces of the system should be kept clean for
hygienic reasons and the applicator tips should be cleaned between sessions to
enable effective treatment.
Warning
Maintenance by the operator should be performed only when the system is shut
down and disconnected from the mains power source. Performing maintenance
procedures with the system turned on may be hazardous to the operator and/or
destructive to the system.
7.2 Periodic Service
The system should be periodically inspected and maintained to keep it in peak
operating condition.
The following routine service operations should be performed by Alma Lasers
authorized technical personnel every 12 months:
• General system check
• Interior inspection, including cleaning accumulated layers of dust
• Check the radiofrequency output energy
If periodic service is not performed according to schedule (every 12 months) by an
Alma Lasers-authorized service provider, any warranty agreement of the system
and its associated applicators and/or accessories will be invalidated.
Warning
The system generates hazardous voltages within the main console.
The interior of the system may be serviced only by Alma Lasers authorized
technical personnel.