Electromagnetic Compatibility
Opus Operator's Manual A-3
Guidance and manufacturer’s declaration electromagnetic IMMUNITY IEC
60601-1-2 Ed.4
Immunity test
IEC 60601 level
Compliance level
IEC 61000-4-6
Conducted RF
3 Vrms
150 kHz to 80 MHz
6 Vrms in ISM bands (6.765 MHz to 6.795 MHz; 13.553 MHz to
13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70
MHz)
[V] = 3 Vrms
[V] = 6 Vrms
IEC 61000-4-3
Radiated RF
3 V/m
80 MHz to 2.7 GHz
[E] = 3 V/m
Proximity fields
from RF wireless
communications
equipment
385 MHz
27 V/m
450 MHz
28 V/m
710 MHz
9 V/m
745 MHz
780 MHz
810 MHz
28 V/m
870 MHz
930 MHz
1720 MHz
28 V/m
1845 MHz
1970 MHz
2450 MHz
28 V/m
5240 MHz
9 V/m
5500 MHz
5785 MHz