EasyManua.ls Logo

Alpha AS100 Series - Page 100

Alpha AS100 Series
179 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter VI Specified Function Introduction
99
Function Description:
Refer to P03.04 for function description.
Function
Code
Parameter Name
Attribute
Setting
Range
Unit
Factory
Defaults
P03.11
Zero deviation
calibration for analog
input
Immediately
effective
0~1
-
1
Function Description:
Set the auto zero calibration for analog input. This setting is valid when Analog Speed
Control mode or Torque Control mode is set in P00.02.
0: Start the Analog input Auto Zero Calibration (Make sure the actual analog input is
zero prior to start auto zero calibration).
System will automatically check the analog input, and save the measured value in
P03.12. Entering function code P03.12 and pressing button to save the
measured value into EEPROM is needed. Only one auto zero calibration will be
operated during each time when power up.
(After auto zero calibration is done, the set point must be 1, or otherwise exceptions
will occur during power up next time).
1: Analog Auto Zero Calibration Invalid
Function
Code
Parameter
Name
Attribute
Setting Range
Unit
Factory
Defaults
P03.12
Analog input
zero
compensation
value
Immediately
effective
-5.000~+5.000
0.001V
0.01
Function Description:
Set the Analog Input Zero Compensation value. The setting is applying to Analog
Speed Control Mode and Analog Torque Control Mode.
The Zero Compensation value can be acquired by operating the Analog Auto Zero
Calibration in P03.11, and also it can be acquired by manual settings. The displaying of
Zero Compensation value will be voltage style, which displayed three decimal places.

Table of Contents