2 - 1
Lancom 4Portable Flue Gas Analyser
User Guide
2 Dos and Don’ts
DO
Use the analyser in the
upright position.
Make sure there are no loops
in the sample line.
Connect the sample line
before operating.
Makesuretheltersarein
good condition.
Empty the catch-pot after use.
Allow the temperature of the
analyser to stabilise before
use, for example if it has been
stored in an air-conditioned
oceorahotcar.
If you are using the analyser
between locations where the
ambient temperature may
vary, use the recalibrate
function between locations to
optimise performance.
Recalibrateusingcertied
gas mixtures every 6
months. Contact LAND or an
authorised dealer for details.
Ensure that the sample
exhaust is in a well ventilated
area.
DO NOT
Abort the zero calibration –
this will result in incorrect gas
readings.
Operate the analyser on its
side – liquid in the catch-pot
mayoodtheanalyser.
Leave the Lancom 4 operating
unattended for long periods
–thecatch-potmaylland
oodtheanalyser-usethe
“wake and sleep” option
instead.
Use for extended periods
where gas concentrations are
very high.
Attempt to sample gas
without connecting the
sample line - it may cause
damage to the pump.
Operate with the catch-pot
orltersremoved-thiswill
result in incorrect readings.
Touch the probe tube after
use – it could be very hot.
Over-tighten the catch-pot.
Leave the analyser in direct
sunlight.
Use the probe without the
sinterlter.
Use the Lancom 4 in a
connedspacewithout
connecting an exhaust tube
to the sample outlet.