1 - 3
Lancom 4Portable Flue Gas Analyser
User Guide
1.3 Probe Connections
Probe/thermocouple connector
Flue gas port
Flow probe port
Exhaust Port (See Warning below)
1.2 LED Indicators
Standby The analyser is running in ‘Wake & sleep’ mode
On The analyser is powered and operational.
Charge The battery is being charged or the analyser is connect-
ed to an external power supply.
Low Battery Recharge the batteries. See Section 8.1.
Service Carry out routine maintenance. See Maintenance
Schedule. See Section 8.
WARNING
Exhaust Port
-Whenmeasuringuegas,wastegasisemittedfromthe
exhaust port. Flue gas can be toxic. Ensure good ventilation around the
exhaust port. If you are operating the instrument in an enclosed space,
attach an extension to the exhaust port to remove exhaust gas.
Note:
Gasreadingaccuracymaybeaectedifanyback-pressureis
applied to the exhaust port. Therefore, any exhaust piping must have
the largest diameter possible.