11 - 5
Lancom 4Portable Flue Gas Analyser
User Guide
11.3 Maintenance Schedule
Item Check
Sinter lter
702.182
Keepthesinterlterclean.Afteruse,allow
to cool and clean with a wire brush and a mild
detergent.
Replace if it is cracked or corroded.
Donotoperatewiththesinterlterremoved.
Gas pre-lter
703.144
Replace if:
thelterisdirtyordiscoloured;
the plastic casing is cracked or opaque.
*CO compen-
sation lter
703.145
Replace if:
the service light illuminates;
theltergranulesbegintoturnwhite;
the plastic casing is cracked or opaque.
Catch-pot
701.828
Replace if cracked or damaged.
Check that the gasket (319.250) is still in
place inside the hinged section.
Probe O-ring
319.211
Remove the probe from the handle and check
the O-ring. Replace if damaged.
*ResettheltercounterwheneveryoureplacetheCOcompensationlter(SeeSection8.3)
Upgrading Your Analyser
Ifupgradesareavailablefortheinstrumentrmware,thesecanbesentvia
email.
To install the update, connect the instrument to a PC via the USB cable,
CopythesuppliedfoldersandlestotheLancom4.SwitchtheLancom4o
and back on again to complete the update.