EasyManua.ls Logo

Ametek Land LANCOM 4 - Using the High Temperature Probe

Ametek Land LANCOM 4
81 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 - 3
Lancom 4Portable Flue Gas Analyser
User Guide
6.2 Using the High Temperature Probe
TheLANDHighTemperatureProbecanbeusedinuegastemperaturesupto
1400°C (2552°F)
The High Temperature Probe is used in
the same way as the standard probe,
however it cannot be used to measure
fueleciencyasthegastemperatureis
not measured.
CAUTION
The Probe Pipe and Heat Shield become HOT when in use.
See Section 4.3 for a table of entry port diameters.
Assembling the High Temperature Probe
1) Refer to the illustration above. Loosen the sealing cone (2).
2) Slide probe pipe (1) through the sealing cone (2) and seal (3) and
carefully tighten the sealing cone (2).
When replacing a probe pipe, remove
the seal (3) and clear any debris before
replacing the probe pipe.
3) The position of the probe handle (7) can
be adjusted. Loosen the top retaining
screws and reposition the handle.
Tighten the screws securely.
Using the probe
1) Connectthesampleline(5)totheue
gas port on the analyser.
2) From the Main Menu, select Readings.
The high temperature probe is used in the
same way as a standard probe.
1 2 3
4
5
6 7
Probe
Flue Gas
Flow Probe
Exhaust
Connect the sample probe to
the ‘Flue Gas’ port

Table of Contents