Portable Flue Gas Analyser
User Guide
Lancom 4
6 - 2
4) Connecttheowmetertotheexhaustport.
5) Usetheowmeterreadingtodeterminethesampletimefromthe
table.Donotdisconnecttheowmeter.
Flow Rate (l/min) Sample Time (s) Flow rate (l/min) Sample Time (s)
0.4 240 1.5 64
0.5 192 1.6 60
0.6 160 1.7 57
0.7 137 1.8 53
0.8 120 1.9 51
0.9 107 2.0 48
1.0 96 2.1 46
1.1 87 2.2 44
1.2 80 2.3 42
1.3 74 2.4 40
1.4 69 2.5 38
Sample Times for LAND Smoke Probe
Example
Iftheowmeterreads0.7litresperminutesetSMOKEMEASUREMENTto
137 seconds.
The heated smoke probe will operate from fully charged batteries for a
maximum of 20 minutes.
If you intend to use the smoke probe for longer periods use a mains power
supply.
Making the smoke measurement
1) From the main menu, select the Smoke measurement option.
The Wait for the probe to heat then insert the smoke lter
message is displayed.
2) Press enter to continue.
The Smoke measurement countdown screen is displayed for 30
seconds, followed by Remove the smoke lter message.
3) Thesampleowrateshouldremainconstantthroughoutthesmoke
measurementtime.Ifthesampleowratedropssignicantlyremove
thelterpaperandrepeatthetest.
4) Whenthesmokemeasurementiscompleteremovethelterpaper.
5) ComparethelterpaperwiththeSmokeSpotScale.