Portable Flue Gas Analyser
User Guide
Lancom 4
11 - 4
To Check the Battery Power
1) From the Main Menu, select the Diagnostics option.
The Diagnostics screen is displayed.
2) Select the System option.
The Battery status (remaining charge) is displayed at the bottom of the
system diagnostics screen.
Whenfullychargedthebatterywillread100%.Thisissucientfor
about 8 hours use.
When the battery voltage falls below 11.5 Volts the Low Battery light
will illuminate.
NOTE
If the battery voltage falls below 10.5 Volts, the analyser will shut
down.
11.2 The Service Light
If the Service light illuminates, check the diagnostics screen.
Contact LAND or an approved distributor.
1) From the Main Menu, select the Diagnostics option.
The Diagnostics screen is displayed.
2) Select the Faults option.
Fault Report Action
X compression fault Calculation error. Contact LAND
Zero O2 readings following recal Oxygen sensor failed - replace sensor.
Contact LAND
Main battery low Recharge batteries. See 8.1.
Filter change recommended
Changeltersandresettheltercounter.
Clearing Faults
1) From the main menu, select the Maintenance option.
The maintenance screen is displayed.
2) Select the Clear faults option.
The fault log will be cleared and the fault light will go out.