Battery Replacement
CAUTION
Do not short-circuit the battery terminals.
If it is necessary to change the batteries:
1) Disconnect the charger and other connected equipment.
2) Unscrew the side panel screws and
remove both instrument side panels.
3) Disconnectthebatteriesbypullingothe
spade terminals.
4) Press out the batteries and recycle
according to local environmental laws, or
return to Land Instruments for recycling.
The batteries are Sealed Lead Acid type
5) Fit the self adhesive packing foam to the
batteries.
6) Press the batteries into the case.
7) Retthebatteryterminals,ensuringthe
correct polarity:
Red wire Positive terminal
Black wire Negative terminal
8) Rettheinstrumentsidepanels.
9) Connect the power supply and ensure
thechargeLEDashes,thenafterafew
seconds is lit continuously.
10) Switch the instrument ON
11) Leave the instrument to charge fully.
12) Switchtheinstrumento.
The charger can be left connected to
keep the instrument ready for use.
The Low Battery light on the control
panel illuminates when the batteries
need recharging.
Do not remove the batteries. Connect
the analyser charger to the mains power
supply and plug the charger into the
analyser.
The batteries will recharge while the
analyserisswitchedo.Thecharging
light on the control panel will be
illuminated.
A full recharge will take about 8 hours.
Side panel screws
Batteries