14 - 1
Lancom 4Portable Flue Gas Analyser
User Guide
14 Technical Specification
Sensor Detection Limit Full Scale Range Upscale Repeatability Resolution
O
2
0.2 % v/v 0 to 30% v/v ±1 % 0.1% v/v
CO (low) 2 ppm 0 to 6000 ppm ±2 % 0.1 ppm
CO (H
2
compensated) 2 ppm 0 to 4000 ppm ±2 % 0.1 ppm
CO (high) 20 ppm 0 to 10 % v/v ±2 % 0.1 ppm
SO
2
2 ppm 0 to 4000 ppm ±2 % 0.1 ppm
NO 2 ppm 0 to 5000 ppm ±2 % 0.1 ppm
NO
2
2 ppm 0 to 1000 ppm ±2 % 0.1 ppm
H
2
S 2 ppm 0 to 1000 ppm ±2 % 0.1 ppm
CO
2
** 0.2 % v/v 0 to 25 % v/v ±2 % 0.1% v/v
Hydrocarbons (C
X
H
Y
) Application dependent 0 to 4 % v/v ±4 % 0.1% v/v
Flue Gas/Ambient
Temperature
Measured
Draft ± 50 hPa / 20 “ Water Gauge ***
Flow (velocity) 1 to 50 m/s
Fuels:
Wood,LightFuelOil,HeavyfuelOil,NaturalGas,Propane,Coal,4User-denablefuels
Power Supply: 95-265 V a.c. ±10%, 50-60 Hz, 30 Watts. Battery, rechargeable lead-acid (internal).
Typical8houroperation,dependentonoptionstted
Ambient Temperature: -5 °C to 45 °C (+23 °F to 113 °F)
Flue Temperature: Max. 600 °C (1112 °F) standard, Max. 1400 °C (2552 °F) with high-temperature probe
Case: Medium density blended polyethylene
Dimensions: 453 x 120 x 245 mm (17.8 x 4.7 x 9.6 in)
Weight: 6kg (13lb)
Standard Accessories:
Integralwatercatch-potandlters,BatteryChargersupply,Probehandle,Hoseandprobe
pipe (Lengths listed below under options), Data logging
Options: Min of 3 to max 9 gases in total, from a selection of 9 gases
Probe length options - 0.3, 1.0, 1.5, 2.0, 3.0 m/1, 3.3, 5, 6.5, 10 ft
Alternative probes available - refer to Data Sheet Reference PDS198 for details
Hose length options - 3 m/10 ft or 10 m/33 ft
Draft Measurement
Flow Measurement, probe length options - 0.7, 1.2, 2.2, 3.0 m/2.3, 3.9, 7.2, 9.8 ft
Smoke Measurement, probe length options - 0.3, 0.75, 1.0 m/1, 2.4, 3.3 ft
Insight Data Acquisition Software system - Refer to Data Sheet Reference PDS205 for details
Analogueoutputs(8currentloops,4to20mA,independentlycongurable)
Wake and Sleep facility (Semi-continuous monitoring)
Language display options - English, French, German, Italian, Spanish & Chinese