PN 305725901, Rev YF
Alarm Condition / Description / Corrective Action LED Flashes
per Cycle
Flow Out of Tolerance* 5
Sample ow rate too high or too low.
Corrective Action:
Check the Inlet and Outlet pressure. Call AMETEK Service if the problem persists.
Battery Low* 6
Battery needs to be replaced.
Corrective Action:
Call AMETEK Service.
Reference Gas* 7
Analyzer detected problem with Reference gas.
Corrective Action:
Check and/or replace the Dryer. Call AMETEK Service if the problem persists.
Enclosure Temperature 8
Excessive internal temperature.
Corrective Action:
External temperature should be 80 °C or less. Call AMETEK Service.
Moisture Generator Date N/A
Moisture Generator date has expired.
Corrective Action:
Replace the Moisture Generator.
Dryer Alarm 10
Dryer failure Imminent.
Corrective Action:
Replace the Dryer.
Concentration Alarm 11
Moisture concentration is out of user dened limits.
Corrective Action:
Review alarm settings and use the Congurator Software to identify source of error.
The Data Valid contact Opens on all alarms and stays Closed during
normal functions and readings. An Open Data Valid contact indicates
Verication is in process or an alarm condition. The RED LED ashes 11
times if an invalid condition occurs.
4-6 | 3050-DO Moisture Analyzer