PN 305725901, Rev YF
Alarm Condition / Description / Corrective Action LED Flashes
per Cycle
Flow Out of Tolerance* 5
Sample ow rate too high or too low.
Corrective Action:
Check the Inlet and Outlet pressure. Call AMETEK Service if the problem persists.
Battery Low* 6
Battery needs to be replaced.
Corrective Action:
Call AMETEK Service.
Reference Gas* 7
Analyzer detected problem with Reference gas.
Corrective Action:
Check and/or replace the Dryer. Call AMETEK Service if the problem persists.
Enclosure Temperature 8
Excessive internal temperature.
Corrective Action:
External temperature should be 80 °C or less. Call AMETEK Service.
Moisture Generator Date N/A
Moisture Generator date has expired.
Corrective Action:
Replace the Moisture Generator.
Dryer Alarm 10
Dryer failure Imminent.
Corrective Action:
Replace the Dryer.
Concentration Alarm 11
Moisture concentration is out of user dened limits.
Corrective Action:
Review alarm settings and use the Congurator Software to identify source of error.
The Data Valid contact Opens on all alarms and stays Closed during
normal functions and readings. An Open Data Valid contact indicates
Verication is in process or an alarm condition. The RED LED ashes 11
times if an invalid condition occurs.
4-6 | 3050-DO Moisture Analyzer