PN 305725901, Rev YF
Specications
Specications included in this chapter are typical for the ranges listed.
For custom ranges/applications, or consult with your AMETEK repre-
sentative.
3050-DO Moisture Analyzer Specications
Specication Description
Ranges 0.02–100 parts per million by volume (PPMV)
Readings can also be displayed in units of PPMW, lb/mmscf,
and mg/Nm
3
.
Accuracy 0.02 PPMV or ± 10 % of reading, whichever is greater
Reproducibility ± 5 % of reading from 0.2–2500 PPMV
Moisture Generator Value 1 PPMV nominal
Back Pressure Regulator
Requirements (if
purchased)
0–1 BAR (0–15 PSIG) maximum
Sample Inlet Pressure
Requirement
138–24150 kPa (20–3500 PSIG) to pressure reducer
138–345 kPa (20–50 PSIG) maximum, pressure reducer to
analyzer
Exhaust Pressure 0–1 BAR (0–15 PSIG)
Minimum Dierential
Pressure
20 PSIG (Inlet pressure must be greater than 20 PSIG over the
outlet pressure).
Data Isolated 4–20 mA, 100–500 Ω Analog output (software
congurable)
RS-232 or RS-485 serial port, 2 and 4-wire mode
Minimum PC
Requirements for 3050
Congurator Software
Pentium 100
32 MB RAM For Windows XP
Alarms/Alerts Two independent contact closures 60 VDC, 30 VAC, 50 VA
maximum resistive for System Alarm and Data Valid. All are fail-
safe by default.
Alarms are available on RS-485 interface.
Lower Detectable Limit 0.02 PPMV nominal
Sensitivity 0.01 PPMV or 1 % of reading, whichever is greater
Environmental Ambient Temperature: Enclosed sample system:
-20 °C to 45 °C (-4 °F to 113 °F)
Relative Humidity: 90 % RH maximum, non-condensing
Maximum Altitude: 2,000 meters (6,560 feet)
IEC Pollution Degree 2
IEC Installation Category II
Indoor/Outdoor Use
Specications | 5-1