PN 305725901, Rev YF
Specication Description
Sample Flow Requirement 150 SCCM, 1 SLPM bypass ow available for increased response
speed.
Sample (Inlet) Gas
Temperature
0 °C to 100 °C (32 °F to 212 °F)
Optimal results are obtained when inlet gas temperature is
maintained at 60 °C (140 °F).
Heat traced sample lines are recommended.
Physical Dimensions Width: 61.2 cm (24.11 inches)
Height: 64.6 cm (25.42 inches)
Depth: 34.2 cm (13.47 inches)
Weight: 82 kg (180 lb)
Utility Requirements Power Consumption
115 ±10 % VAC, 50/60 Hz, 150 W maximum
230 ±10 % VAC, 50/60 Hz, 150 W maximum
24 VDC 50 W maximum
Instrument Air
5–7 BAR (70–100 PSIG)
3050-DO Moisture Analyzer Approvals and Certications
Design Classications Designed to meet the requirements for General Purpose
Locations/Electrical Safety and the following Hazardous
Location classications:
North American Class I, Division 2, Groups A,B,C,D T4
North American Class I, Division 1, Groups B,C,D T6
Electromagnetic Compatibility Directive:
EN61326-1 Industrial
Low Voltage Directive: EN61010-1
Pressure Equipment Directive
Russian Gosstandart Pattern Approval
3050-DO Moisture Analyzer Marking
5-2 | 3050-DO Moisture Analyzer