Specifications
For a complete listing of all analyzer specifications, refer to the analyzer User Manual.
Utility Requirements
Power Consumption
Without Oven Heater
< 300 W maximum start-up (from a cold start) with continuous average, depending on ambient
temperature.
With Oven Heater
< 700 W maximum start-up (from a cold start) with continuous average, depending on ambient
temperature.
Supply Voltage
120 VAC (105–132 VAC), 47–63 Hz OR
240 VAC (209–264 VAC), 47–63 Hz
Instrument Air
The aspirator medium must be transported in 316 stainless steel.
Minimum pressure:
490–700 kPag (70–100 psig / 4.8–6.9 barg)
Air Quality:
As per ANSI/ISA-S7.0.01 (1996) Quality Standard for Instrument Air.
Outputs
256 x 64 pixel vacuum-fluorescent display with multilingual capability.
Four (4) isolated analog outputs, configurable as 0–20 mA or 4–20 mA(two are standard; two are
optional).
Up to eight (8) relays (NO, contact rating 100 VA, 240 V maximum), which indicate the opera-
tional status of the analyzer. Each relay provides one SPST (Form A) dry (potential free) contact.
Two Form C loss of purge alarm contacts rated 240Vac 5A max
RS485 serial port, isolated.
RS232 serial port, non-isolated.
Essential Health & Safety Requirements | 13