Fast Ethernet (IEE802.3).
Modbus TCP (Supports single Modbus Client).
Ambient Temperature Limits
-20 °C ≤ Ta ≤ 50 °C , without external heating or cooling.
Sample System Limits
Sample Pressure
Up to 10,000 kPag (1,450 psig / 100 barg) can be accommodated, depend ing on the application.
See product markings or customer-specific drawings in the “Supplemental Information” tab of
this manual.
Oven/Sample System Enclosure Temperature
For applications with a heated sample system, the Oven will operate at temperatures up to 150
°C (302 °F).
Approvals and Certifications
The IPS-4 Analyzer is certified with regard to electrical safety for Pollution Degree 2 (normally
nonconductive environmental pollution occurs with occasional condensation) and Installation
Category II (local level transients, less than those found at power distribution level). Mains sup-
ply voltage fluctuations are not to exceed 10 percent of the nominal supply voltage.
Approvals and certifications for the Zone1/Division1 Model include:
NEC/CEC: Class I Div 1 Groups A, B, C, D, Class II Div 1
Groups F & G, and Class III
Class I Zone 1 AEx d px IIC T3, Class I Zone 1
Ex d px IIC T3
ATEX: II 2 G ATEX Ex db pxb IIC T3 Gb IP65
NOTE - models with the 4660 sample system flowmeter installed are reduced to gas groups
BCD and IIB+H2.
NOTE
14 | IPS-4 Zone1/Analyzer