Figure 5-7
Prepare Load Screen
When the load is properly connected press ENTER. At this time the supply will output a
small voltage and request that you tell it exactly what that voltage is. The display will be
as shown in Figure 5-8.
Figure 5-8
Voltage Calibration 1
In this figure the number 230 is for example only and will vary with each supply. The
“OK” means the value being read internally is within an acceptable range and the
calibration may proceed. At this time read the output voltage using the volt meter, and
using the numeric keypad enter the correct voltage to five places. Press ENTER when
the correct voltage has been input. The next voltage calibration step is to take a reading
at a high output voltage. After the previous ENTER is pressed the warning screen
shown in Figure 5-9 will be displayed. On high output voltage supplies the voltage at the
output of the supply may be hazardous and proper precautions should be taken.
Figure 5-9
High Voltage Warning Screen
M550004-01
5-4