A-38 | Thermox WDG-V Series Analyzers
Methane Detector
This screen displays the information you need to determine the health of
the methane detector and to determine if and when it needs to be re-
placed.
Figure42.Diagnostics_MethaneDetectorscreen.
Detector Age in days
Sensitivity mV/%
Active Element Voltage Volts
Reference Element Voltage Volts
I/O - Inputs and Outputs
Diagnostics Mode is turned on when changes are made on this menu
so that the system does not accidentally trigger an auto calibration
while troubleshooting.
Figure43.Diagnostics_I/Oscreen.