A-26 | Thermox WDG-V Series Analyzers
2. When you have completed this screen, click on the SET button to
enter the values.
The automatic calibration has now been set and the system will carry out
the calibration per the schedule entered. When all gases have calibrated,
the screen displays the Diagnostic Results.
On-Demand Auto Calibration
This type of calibration requires the operator to physically click on the
START button to start the calibration. The real-time sequence of calibra-
tion events is displayed and requires input from the user before each step
of the calibration continues.
Use this method for remote calibration and for immediate auto calibra-
tion.
This is an operator inter-active process in which the operator must
respond to prompts on the screen during the auto calibration.
1. Click on the Calibration button.
The Set Calibration Type screen appears (Figure 25). From the drop-
down menu, select the type of calibration you want to perform:
Manual, Remote or Auto. Click NEXT.
Figure25.Calibration_SetCalibrationTypescreen.