Chapter 3: HOW TO TAKE CARE OF THE SYSTEM
IMPORTANT! For the daily or weekly disinfection of the surfaces of the Genie
®
II Mk2, the
touch screen and the wire the disinfectant Meliseptol
®
Wipes sensitive from B. Braun
Melsungen AG or something comparable is suitable. Due to the use of those wipes the
chance of bringing liquids to the optics is minimized.
IMPORTANT! It is absolutely necessary to prevent the penetration of liquids into the
optics. Therefrom it is not allowed to pipet with inserted tubes. All pipetting steps have to be
made outside the Genie
®
Mk2 device. In case of liquid entry into the optical unit, contact
AmplexDiagnostics GmbH to schedule maintenance or repair of the Genie
®
II Mk2 device.
IMPORTANT! Any unauthorized modification of the device as well as damages caused by
wrong handling entails the invalidity of the warranty.
IMPORTANT! Genie
®
II Mk2
uses a proprietary tube strip that maximises optical and
thermal efficiencies. Other tubes and strips will not fit.
IMPORTANT! Forcing non-supported consumables will cause damage to the instrument
and invalidate the warranty.
IMPORTANT! The shape of the tubes is such that they will only fit in one way round. The
locating pins on the block have corresponding holes in the strips.
IMPORTANT! Gently lift the lids upwards. Close the lids by lowering gently. Care must be
taken to ensure that objects are not obstructing the lids when trying to close it and under no
circumstances should the lids be forced open or closed.
IMPORTANT! To maintain the battery life as long as possible, disconnect the device from
time to time and leave the battery almost completely discharge during operation (battery
symbol red). Only then reconnect the device to the mains.
IMPORTANT! The battery only charges when the device is switched on via the main
switch on the back. To save energy, switch the device to standby mode (page 35) or turn it off
completely when not in use.