SITE PREPARATION:
The laboratory bench should be level and stable. The device should be placed centrally on the lab
bench and the surfaces surrounding the device must be clear of obstructions at all times. In the event
of an emergency or under abnormal operating conditions the location should provide, at any time,
enough space to allow easy disconnection of the device.
Care must be taken not to unduly restrict the air at the front of the device and the outlet vents at the
rear. Restricting airflow may impede operation and could affect performance.
Electrical points should be close to the device to avoid injury from trailing wires.
It is recommended that the device is kept away from sinks and other wet areas. Genie
®
II Mk2 is an
electrical device and care should be taken not to operate if there is a risk of water damage.
The Genie
®
II Mk2 device is a highly sensitive and precise optical instrument. Results can be affected
by vibration if the instrument is used too close to a vibrating machine.
IMPORTANT! If a Genie
®
II Mk2 device is kept in a very cold environment, the battery will not
begin charging until the internal temperature has reached 15°C.