Figure 52. Mode Tab
Available Options
Table 4. PPG Mode Configuration Options
Configures PPG raw mode. HR/SpO
2
algorithm outputs are
disabled. PPG settings are controlled manually through the
‘PPG Settings’ and ‘PPG Measurement Settings’ tabs.
Enables HR/SpO
2
algorithm outputs in the ‘Algorithm Hub’
operating mode. See the Algorithm Operating Modes section
for more details.
Enables AGC (see ‘Control Select’).
Automatic gain control = AGC. This mode offers some more
control than AEC. In this mode, the algorithm hub is given a
target PD current, and it adapts the LED selection and driver
currents based on what it predicts it needs to hit that target.
Sample rate and DAC offsets can be manually set in this
mode.
With AGC enabled: This is the fixed PD current target the
algorithm hub always tries to hit. It adjusts LED driver currents
as necessary.