Re marks: Places a sin gle AS CII char ac ter '1' on the GPIB out put queue when the
con di tions for the *OPC com mand are met. An operation is complete
when all input messages before the command have been completed and
any responses have been read out of the output buffer.
Ex am ple: RGH A, 1; RGH B, 2; *OPC?
Re turns a ‘1’ on the GPIB out put when it has fin ished set ting the range
hold com mands.
*RCL Re call a stored setup
SYS TEM
Syn tax: *RCL <val>
val: 1 to 10
Re marks: The ML2400A Se ries can store up to 10 in stru ment con fig u ra tions for
con ve nient re call. The con fig u ra tion pa ram e ters stored are:
Sen sor Setup, Chan nel Setup, and Trig ger Setup.
This com mand sets the ML2400A Se ries to a con fig u ra tion pre vi ously
stored in mem ory lo ca tions 1 through 10. Trying to re call a setup that
has not been saved will set the ex e cu tion bit in the event reg is ter (EXE
in ESR). When re call ing a setup in volves the read out mode chang ing be -
tween STAN DARD and CDMA, the in stru ment will re boot.
Re lated
Com mands: *SAV
*RST Re set De vice
GPIB 488.2
Syn tax: *RST
Re marks: Re sets the ML2400A Se ries to the de fault con fig u ra tion (see Ap pen dix
A, Sec tion A-3 ). Off set ta bles are not cleared. The GPIB ad dress and em -
u la tion set tings are not changed, and the in put queue, out put queue,
and sta tus reg is ters on the GPIB are not cleared. The read out mode will
be set to STAN DARD. This com mand pro duces the same re sult as the
front panel key se quence
Sys tem|Setup|PRESET|RE SET .
*SAV Save con figu ra tion
SYS TEM
Syn tax: *SAV <val>
val: 1 to 10
ML2400A OM 6-15
GPIB OPERATION ML24XXA NATIVE COMMANDS