EasyManua.ls Logo

Anritsu ML2400A Series - Page 130

Anritsu ML2400A Series
275 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
<Delta_power> and <Delta_time> are ab so lute val ues. <Av er age> is
only pres ent when be tween cur sor av er ag ing is turned ON with the
GRAVG com mand.
If no data is avail able, that is, a sen sor is not fit ted, the pro file is not
trig gered, or the Power vs. Time graph has not reached to the cur sor, the
out put for the rel e vant read out value is 999 out put as 9.99e2.
Ex am ple re sult strings might look like:
GRDRQ 1, –10.000, –5.000, 5, 1E- 03 (cur sor av er age off)
GRDRQ 1, –10.000, –5.000, 5, 1E- 03, –7.5 (cur sor av er age on)
Re lated
Com mands:
GRAVG , GRDATA
GRFS Pro file Ref er ence line state
PRO FILE SETUP
Syn tax: GRFS <state>
state: ON or OFF
Re marks: Turn the pro file ref er ence line ON or OFF. The pro file ref er ence line is
cen tered be tween the top and bot tom of the dis play.
Query: GRFS?
Re turned
String: GRFS <state>
GRMD Pro file, Power vs. Time and Source Sweep Mode Chan nel
Se lec tion
PRO FILE SETUP
Syn tax: GRMD <c>
c: 1 or 2
Re marks: Se lects the chan nel dis played on the Pro file, Power vs. Time and Source
Sweep graphs.
Query: GRMD?
Re turned
String: GRMD <c>
GRPIX Pro file type
PRO FILE SETUP
Syn tax: GRPIX <mode>
ML2400A OM 6-43
GPIB OPERATION ML24XXA NATIVE COMMANDS

Table of Contents

Related product manuals