OBVSP BNC ana log out put stop volt age scale BNC
Syn tax: OBVSP <port>,<val>
port: 1 or 2
val: –5.00 to +5.00 Volts
Re marks: Sets up the stop value for the volt age out put in an a log out put mode. At -
tempting to set the start value to a volt age greater than the stop value,
or the stop value lower than the start value, will re sult in an ex e cu tion
er ror.
Query: OBVSP? <port>
Re turned
String: OBVSP <port>,<val>
OBVST BNC ana log out put start volt age scale
BNC
Syn tax: OBVST <port>,<val>
port: 1 or 2
val: –5.00 to +5.00 Volts
Re marks: Sets up the start value for the volt age out put in an a log out put mode. At -
tempting to set the start value to a volt age greater than the stop value,
or the stop value lower than the start value, will re sult in an ex e cu tion
er ror.
Query: OBVST? <port>
Re turned
String: OBVST <port>,<val>
OBZL BNC RF blanking output level when zeroing
BNC
Syn tax: OBZL <level>
level: HIGH (TTL high)
LOW (TTL low)
Re marks: Sets the TTL level of the BNC RF blank ing out put.
Query: OBZL?
Re turned
String: OBZL <level>
ML2400A OM 6-61