Procedure
To carry out the Self Verication test:
1.
In the Home screen, touch Seings.
2.
In the Seings screen, touch Maintenance & Services.
3.
In the Maintenance & Seings screen, touch Self Verication Test.
4.
In the Self Verication Test screen, touch Start Test to begin testing or Cancel to
exit the screen
The test takes about 10 minutes. Once the test is completed the test results will be
displayed in the form of a report.
Note: Contact your service representative in case of block failure.
5.
Touch Export to save the test report to a USB drive or Close to exit the screen.
Order kits and replacement parts
Refer to Appendix B, “Parts and materials“to obtain part numbers for kits and
replacement parts.