EasyManuals Logo

Applied Biosystems SimpliAmp Thermal Cycler User Manual

Applied Biosystems SimpliAmp Thermal Cycler
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #63 background imageLoading...
Page #63 background image
Instrument specification and layout
This chapter lists the instrument specications for the SimpliAmp
Thermal Cycler
using the 96-Well Sample Block.
Site requirements
The SimpliAmp
Thermal Cycler is for indoor use. Ensure that the installation site:
Meets the spatial and weight requirements (see “System specications“ on
page 65)
Meets environmental requirements (see “Environmental requirements“ on
page 63)
Is within 1 m (3 ft.) of an AC power source receptacle
Is away from water
Environmental requirements
Ensure that the installation site is maintained under the following conditions:
Table 8 Temperature and humidity requirements
Condition
Acceptable range
Temperature 15 to 30°C (59 to 86°F)
Humidity 15 to 80% relative humidity, non-
condensing
Avoid placing the instrument adjacent to heaters, cooling ducts, or in direct sunlight.
Fluctuations between day and night temperatures can cause system instability. Place
away from any equipment that vibrates, such as a refrigerator or centrifuge.
D
Temperature and
humidity
requirements
SimpliAmp
Thermal Cycler User Guide
63

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems SimpliAmp Thermal Cycler and is the answer not in the manual?

Applied Biosystems SimpliAmp Thermal Cycler Specifications

General IconGeneral
BrandApplied Biosystems
ModelSimpliAmp Thermal Cycler
CategoryBicycle Accessories
LanguageEnglish