The SimpliAmp
Thermal Cycler has a Pollution Degree rating of 2. It may be
installed in an environment that has non-conductive pollutants only, such as dust
particles or wood chips. Typical environments with a Pollution Degree II rating are
laboratories and sales and commercial areas.
The safety of SimpliAmp
Thermal Cycler use was tested for altitudes up to 6000 ft.
Technical specifications
The following table displays the system specications for the SimpliAmp
Thermal
Cycler.
Table 9 SimpliAmp
Thermal Cycler technical specifications:
Feature
Specification
Maximum block ramp rate
[1]
4°C/sec
Maximum sample ramp rate
[1]
3°C/sec
Temperature accuracy ±0.25°C (35 to 99.9°C)
Temperature range for
protocol run
0°C to 100.0°C
Temperature non-uniformity
[2]
<0.5°C
Dimensions Height: 21 cm (8.27 in)
Width: 24.0 cm (9.45 in)
Depth: 46 cm (18.11 in)
Weight 8.5 kg (18.3 lb)
PCR volume range Supported: 10 - 100 µL
Settable: 1 - 100 µL
Instrument memory USB, on-board
Display interface 8" color TFT LCD
Power 100-240 V, 50-60 Hz Max: 600 W
VeriFlex
blocks 3 VeriFlex
zones
Supported 10°C (5°C zone-to-zone)
Settable
[3]
20°C (10°C zone-to-zone)
Ambient humidity
15% - 80% Relative Humidity, non-condensing
(acceptable range)
Ambient operating
temperature
15°C to 30°C (acceptable range)
Pollution
Altitude
Appendix D Instrument specification and layout