[1]
Reaction volume @ 1 µL
[2]
30 seconds after clock start
[3]
Temperature accuracy ±0.5°C and Temperature non-uniformity < 0.75°C
Table 10 Power and communication ports
Port Description
AC power cable port
10/100 Fast Ethernet port for connecting to a network
USB v2.0 port for connecting to an external network drive, jump
drive, or other USB storage device
RS232 serial communication port for connecting the probe for the
Temperature Verification Kit
USB v2.0 port for connecting to USB-enabled Wi-Fi Card
System specifications
Table 11 Physical dimensions, weight, and power consumption
Parameter
Instrument
footprint
Recommended
clearance
Total
Height 21 cm (8.27 in) 24 cm (9.44 in)
[1]
45 cm (17.71 in)
Width 24.0 cm (9.45 in) 0 cm (0 in) 24.0 cm (9.45 in)
Depth 46 cm (18.11 in) 23.0 cm (9 in)
[2]
69 cm (27.1 in)
Weight 8.5 kg (18.3 lb)
Power
consumption
600 W - Maximum consumption
164 W - Average during thermal cycling
30 W - During sleep/ standby
[1]
With the heated cover open
[2]
At the rear of the instrument to ensure adequate airflow and cooling
Appendix D
Instrument specification and layout
System specifications
D
SimpliAmp
™
Thermal Cycler User Guide
65