Power off the ViiA
7 System
Perform an instrument self test
Use the touchscreen to perform a self test of the ViiA
7 System. After the self test is complete, the instrument generates a detailed
summary of the results. You can save the summary to a USB drive and email them to Applied Biosystems Technical Support.
1
Start the instrument Touch the ViiA
7 System touchscreen to awaken it, then press .
2
Run the test a. In the Main Menu, touch Tools, then touch Run Self Test.
b. In the Self Test screen, touch Start Self Test, then wait for the test to complete.
3
(Optional) Save results to
a USB drive
a. Plug a USB drive into the USB port on the right side
of the touchscreen.
b. Touch Save to USB.
IMPORTANT! Do not remove the USB drive from the
ViiA
7 System until you are instructed to do so.
c. When the ViiA
7 System finishes writing the results
to the USB drive, touch OK, then remove the USB
drive.
4
Return to the Main Menu Touch .
Power off the ViiA
7 System
The Applied Biosystems ViiA
7 Real-Time PCR System operates in low-power mode when not in use; however, the ViiA
7 System can
be powered off completely so that the components draw no power.
1
Power off the ViiA
7
System
a. If the ViiA
7 System touchscreen is not blank, touch to place the ViiA
7 System into stand-
by mode.
b. Toggle the power button on the rear of the ViiA
7 System.
2
Power off the ViiA
7
System computer
a. In the desktop, select StartShut Down.
b. In the Shut Down Windows dialog box, select Shut Down, then click OK.
c. (If necessary) Power off the monitor.
For Research Use Only. Not intended for any animal or human therapeutic or diagnostic use.
LIMITED USE LABEL LICENSE NO. 358: RESEARCH USE ONLY
The purchase of this product conveys to the purchaser the limited, non-transferable right to use the purchased amount of the product only to perform internal research for the sole benefit of
the purchaser. No right to resell this product or any of its components is conveyed expressly, by implication, or by estoppel. This product is for internal research purposes only and is not for
use in commercial applications of any kind, including, without limitation, quality control and commercial services such as reporting the results of purchaser’s activities for a fee or other form
of consideration. For information on obtaining additional rights, please contact outlicensing@lifetech.com or Out Licensing, Life Technologies, 5791 Van Allen Way, Carlsbad, California 92008.
NOTICE TO PURCHASER: LIMITED LICENSE
This instrument is licensed for use under patents owned by the University of Utah Research Foundation and licensed to Idaho Technology, Inc. No right is conveyed expressly by implication or
by estoppel under any other patent claim.
© 2010 Life Technologies Corporation. All rights reserved. The trademarks mentioned herein are the property of Life Technologies Corporation or
their respective owners. TaqMan is a registered trademark of Roche Molecular Systems, Inc.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems ViiA 7 Real and is the answer not in the manual?

Applied Biosystems ViiA 7 Real Specifications

General IconGeneral
BrandApplied Biosystems
ModelViiA 7 Real
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals