2.8 HbA1c Calibration
HA-8190V 2-43
8 Check the measurement results.
“Measuring ...”
Dummy samples and standard solutions will be measured in
that order.
“Results”
The obtained measurement results will be displayed on the screen and printed out.
See “2.9. Displayed and Printed Reports” on page 2-44.
9 Once the calibration is complete
a Check that the calibration rack is not moving, and remove the rack from the rack unloading side of the sampler.
2.8.2 Setting the Calibration Coefficients
a On the standby screen, touch , <1. Calibration menu> and <4. Calibration coefficients setting> in that
order.
b Enter <a> as the gradient and <b> as the intercept of the
calibration coefficients.
c Touch .
The message “Save setting changes?” will appear.
d Touch .
This saves your entries and will return you to the standby screen.
REFERENCE:
About stirring:
Only dummy samples in sample tubes are stirred. The
standard solutions are not stirred.
Measuring dummy samples
Measuring standard solutions