Chapter 4 Maintenance
HA-8190V 4-36
4.3.6 Cleaning the Dilution Container
Wash the dilution container unit and dilution container cover once a month to remove dirt and remnants of the rubber
cap. Inaccurate measurement results may be obtained if the dilution container is dirty.
Prepare: Purified water, cotton swabs, gauze
and protective gloves
1 Prepare for cleaning.
a On the standby screen, touch , <7. Maintenance menu> and <3. Dilution container> in that order.
b Touch .
• The piercing nozzle will move to the rear to make it easy to access the dilution container unit. Liquid will be drained
from the dilution container.
c Wait for the screen shown on the right to appear.
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard used cleaning tools and protective gloves in accordance with local regulations for
biohazardous waste.