2.6 STAT Measurement
HA-8190V 2-27
2.6.1 Measuring a Sample During Normal Measurement
A single sample can be measured by using the STAT port during continuous measurements such as normal
measurement. However, STAT measurements cannot be performed during automatic calibration.
1 Prepare a sample for STAT measurement.
a Prepare a sample in a sample tube or sample cup.
The volume of sample as shown on the right is required for
measurement.
2 Set the STAT measurement conditions.
a Press during normal measurement.
Min. 10 mm
above the
bottom
Sample cup
Min. 400 µL
Sample tube
Min. 1 mL
Whole blood sample
Between
10 and
20 mm
Sample cup
Min. 400 µL
Sample tube
Min. 1 mL
Hemolysis sample
IMPORTANT:
When using sample tubes to measure hemolysis sample, be
sure to prepare the volume of sample as shown on the right.
Inaccurate measurement results will be obtained if there is
too much sample.