3.9 Diagnosis
HA-8190V 3-37
3.9.5 Testing HbA1c Reproducibility (Whole Blood Sample)
Run a reproducibility test following the instructions described below if the reproducibility of HbA1c measurement
results seems to have been reduced in anemia or non-anemia whole blood sample measurements. This test repeatedly
measures the same whole blood sample and displays statistical information (average, R, S.D. and C.V.) from those
results. Since all measurements are performed using sample from a single sample tube, there is no need to divide the
sample into multiple sample tubes before setting them.
Prepare: • For whole blood sample test:
Whole blood sample, sample tube (
1),
normal rack (or start rack for loop transportation) and protective gloves
• For anemia sample test:
Anemia sample, sample tube ( 1), anemia rack and protective gloves
1 Prepare sample.
a On the standby screen, touch , <8. Diagnosis menu>, <4. Reproducibility menu> and <1. Whole sample
meas.> in that order.
The following message will appear: “Set the sample tube in port 1 of the normal rack and load the rack onto the
sampler.”.
b Prepare sample in a sample tube.
Make sure the sample tube is capped.
c Set the sample tube in port 1 of the specified sample rack.
For sample tubes of about 13 mm in diameter, set a port guide in
the port before setting the tube.
d Load the sample rack onto the sampler.
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard used samples and protective gloves in accordance with local regulations for
biohazardous waste.
Performing measurements with uncapped tubes may cause sample to splatter inside the
instrument while spinning, possibly jeopardizing subsequent measurements. It may also cause
infection to the user or other individuals by pathogenic microbes.
NOTE:
The required sample volume (whole blood) is shown below. Add a sufficient volume of sample to the sample
tube to perform the set number of measurements.
(Required sample volume) = 8 L x (Set number of measurements) + 1 mL
MENU
Sample tube with whole blood
12345678910

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray HA-8190V and is the answer not in the manual?

Arkray HA-8190V Specifications

General IconGeneral
BrandArkray
ModelHA-8190V
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals