EasyManua.ls Logo

Arkray HA-8190V - STAT Measurement

Arkray HA-8190V
288 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 Measurement
HA-8190V 2-26
2.6 STAT Measurement
A single sample can be measured by setting it in the STAT port. The STAT port is convenient for interrupting normal
measurements to measure an urgent sample or to measure only one sample.
Prepare: Sample container
and adapter described below,
DILUENT90 (for hemolysis samples) and protective gloves
Sample containers and adapters
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard used samples and protective gloves in accordance with local regulations for
biohazardous waste.
IMPORTANT:
Samples are not stirred in STAT measurements. If blood has sedimented due to centrifuging, invert the sample
tube to mix the sample before performing measurement. Otherwise, inaccurate measurement results will be
obtained.
REFERENCE:
Use sample cups when small volume of sample is available.
Sample tube
(About 13 mm diameter)
Conventional adapter (gray)
Sample tube
(About 15 mm diameter)
Sample cup
Conventional adapter (gray)

Table of Contents

Related product manuals