HA-8190V 1-41
1.6 Relocation
1.6.4 Disconnecting the Tubes, Sensor Cord and Cables
Remove the bottle for liquid waste, hand-held barcode reader and external device from the instrument.
Prepare: Protective gloves
1 Remove the bottle for liquid waste.
a Disconnect the drain tubes from joints “D1” and “D2” on the rear panel.
b When using the optional liquid waste bottle, disconnect the fluid level detection sensor cord from the DRAIN
terminal on the rear panel.
2 Remove the hand-held barcode reader and external device.
When the hand-held barcode reader and/or external device are connected to the instrument, disconnect their cables
from the B.C.R and/or DATA OUT terminals on the rear panel.
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard used protective gloves and liquid waste in accordance with local regulations for
biohazardous waste.