DrillSaw Highspeed 200™ User’s Guide
DFU-0225-2 Rev. 0 Page 18 of 58
3
3
.
.
0
0
Technical Specifications
3.1 Console AR-200C
Table 8 Console (AR-200C) Specifications
Width
256 mm (10 in.)
Height
109 mm (4.3 in.)
Depth
305 mm (12 in.)
Weight
7.0 kg (15.4 lbs.)
Maximum flow rate
90 mL/minute
Power supply
110-130 V / 220-240V, 50/60 Hz, 0.1-1.8A / 0.1-0.9A
allowed voltage
fluctuation:
+/- 10%
Mains fuse
2 x 250V – T1.6AH
Power:
max 180 VA
Cleaning
Refer to Section 6.0 Cleaning and Disinfecting
Sterilization
No
Pollution level
2
Overvoltage category
II
Altitude
Up to 3000 m [9842 ft.] above sea level
Protection
IP22
Table 9 Ambient Conditions for Operation
Temperature
10° to 40 °C (50° to 104 °F)
Relative Humidity
15% to 80% (relative), non-condensing
Table 10 Ambient Conditions for Storage (in shipping packaging)
Temperature
-40° to +70°C (-40° to 158°F)
Relative Humidity
8% to 80% (relative), non-condensing
Atmospheric Pressure
700 hPa 1060 hPa
In the event of a power failure, or if the AR-200C is switched off, or when
switching between programs, the last set of values are saved and re-activated
when the device is switched on again.
A total system failure does not constitute a critical fault. Simply switch the device
off and on again.
In the event of interruption/restoration of power, the device may lose settings or
functions.
In the case of a display failure, the device must be taken out of service and
returned to the service center.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arthrex Highspeed 200 and is the answer not in the manual?

Arthrex Highspeed 200 Specifications

General IconGeneral
BrandArthrex
ModelHighspeed 200
CategorySaw