EasyManuals Logo

Arthrex Highspeed 200 User Manual

Default Icon
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #46 background imageLoading...
Page #46 background image
DrillSaw Highspeed 200™ User’s Guide
DFU-0225-2 Rev. 0 Page 40 of 58
6
6
.
.
0
0
Cleaning and Disinfecting
Follow your country-specific directives, standards and guidelines for cleaning,
disinfection and sterilization.
Always wear proper PPE (Personal Protective Equipment) such as: gloves, eye
protection, face shields, gowns, etc.
To make cleaning and disinfecting easier, always clean and disinfect the Arthrex
DrillSaw HighSpeed 200 System immediately after use; do not allow bodily fluids and
tissue to dry on any of the system components.
Only use disinfectants that have no protein fixing effects.
6.1 Preliminary Cleaning
When properly performed, cleaning, disinfection and/or sterilization do not
compromise the mechanical performance of these instruments. These
instruments are used with or on patients who may harbor, either recognized and
unrecognized infections. To prevent the spread of infection, all reusable
instruments must be thoroughly cleaned, disinfected, and sterilized after being
used on each patient.
Note: No assembly/disassembly of these instruments is required unless stated on
the labeling, directions for use, or literature assembly instructions (LAI)
pertaining to cleaning, disinfection, and sterilization.
1. Devices that require disassembly should be disassembled prior to
cleaning
2. Remove dried-on soil from devices, especially in areas such as joints and
crevices, prior to washing. Clean the surfaces with a sponge or brush
under cold running water until visible contamination has been removed.
3. Check instruments for visible soil. Repeat pre-cleaning if soil is visible
and re-inspect.
4. After the completion of preliminary cleaning, the end user should
proceed to Machine (Automatic) Cleaning and Thermal Disinfection.
6.2 Console (AR-200C)
The AR-200C console is provided non-sterile and must not be sterilized.
The AR-200C console can be cleaned/disinfected using a cloth and commercially
available surfactants or surface disinfectants only.
The AR-200C must not be submersed in any liquid. When cleaning the display,
use a microfiber cloth and gently wipe in a circular motion.
The front panel of the console and the foot control are sealed and may be wiped
clean.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arthrex Highspeed 200 and is the answer not in the manual?

Arthrex Highspeed 200 Specifications

General IconGeneral
BrandArthrex
ModelHighspeed 200
CategorySaw
LanguageEnglish