DrillSaw Highspeed 200™ User’s Guide
DFU-0225-2 Rev. 0 Page 45 of 58
Steam Sterilization Minimum Parameters:
STERILIZATION PARAMETERS: FOR THE USA ONLY:
Exposure
Temperature
Exposure Time Drying Time
Gravity
Displacement
Steam
Sterilization
Cycle
121°C (250°F) 30 Minutes 15 to 30 Minutes
132°C (270°F) 15 Minutes 15 to 30 Minutes
135°C (275°F) 10 Minutes 30 Minutes
Pre-vacuum Cycle
132°C (270°F) 4 Minutes 20 to 30 Minutes
135°C (275°F) 3 Minutes 16 Minutes
STERILIZATION PARAMETERS: FOR OUTSIDE THE USA ONLY:
Exposure
Temperature
Exposure Time Drying Time
Gravity-
Displacement
Steam
Sterilization
Cycle
132°C – 135°C
18 Minutes
15 to 30 Minutes
(270°F – 275°F)
121°C (250°F) 30 Minutes 15 to 30 Minutes
Pre-vacuum Cycle
132°C - 135°C
4 Minutes
20 to 30 Minutes
(270°F - 275°F)
WARNING!
After autoclaving, the accessory devices are VERY HOT. Handle with care to avoid
burns.
Allow to cool down to ambient temperature before using.