DrillSaw Highspeed 200™ User’s Guide
DFU-0225-2 Rev. 0 Page 55 of 58
Table 24 Guidance and Manufacturer's Statement - Electromagnetic Immunity
The AR-200 DrillSaw Highspeed 200 is intended for use in the electromagnetic environment
specified below. The customer or the user of the DrillSaw Highspeed 200 should ensure that it is
used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output
lines
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips,
short
interruptions,
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5 % Uτ
(>95 % dip in Uτ)
for 0.5/1/250
cycle
70 % Uτ
(30 % dip in Uτ)
for 0,5/1/5/10/25
cycles
<5 %
(>95 % dip in )
for 0.5/1/250 cycle
70 %
(30 % dip in )
for 0,5/1/5/10/25
cycles
Mains power quality should be
that of a typical commercial or
hospital environment. If the
DrillSaw Highspeed 200 requires
continued operation during power
mains interruptions, it is
recommended that the DrillSaw
Highspeed 200 be powered from
an uninterruptible power supply.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
100 A/m
100
A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
Note: UT is the AC mains voltage prior to application of the test level.