9. Prior to Installation
The place (bench, table or stand) where the incubator will be set up MUST be steady, firm and
leveled. Make sure the surface supports the weight of the incubator in operational conditions.
To move the incubator, two or more people are needed.
Please keep 10 inches or more of clearance in the back and both sides of the incubator to avoid
heat accumulation.
Avoid placing an incubator where cold air drifts directly to the incubator such as air conditioning or
air vent. It may cause condensation in the chamber.
All the surfaces of the chamber should be cleaned and disinfected before operation.
The incubator should be placed in a clean room environment.
When placing the drawer type incubator on the
optional base stool (P/N: AD-3100STD) balance the
level of incubator adjusting the height of the feet
located at each bottom corner of the incubator and
secure it to the shelf using M6 screw bolts included
in the base stool package. Drawer Type incubator
has dedicated threaded holes on the bottom surface
where indicated by red circles in the right diagram.
Power Source
110, 115, 220, 230 and 240V are available. Please double check for the appropriate power voltage
suitable for the drawer type incubator you have purchased. The power source must be individually
assigned to the unit; we do not recommend multiple connections for any circumstance. The product must
be properly grounded to the earth to avoid electric shock when short circuit.
Connect the incubator to a grounded dedicated circuit only. The power cord connector is the
main disconnect device for the incubator. Position the unit so that it can be easily disconnected.
When connecting the plug of the power cord into the outlet, check the polarity of the plug of the
power cord and the outlet. Connect the outlet of L(Live) pole and L(Live) pole of the power cord
correctly. Incorrect connection can cause electric shock.
Please keep 10 inches or more of clearance in the back and both sides of the incubator/the
dedicated pedestal to avoid heat accumulation.
Do not ground on gas pipe, telephone wire and/or conductor rod.