AD-3100 Drawer Type Incubator Operation Manual Rev. 4
ASTEC CO., LTD.
30
24. Troubleshootings
Problem
Probable Causes:
Unit does not power on.
IS THE POWER CABLE SECURELY PLUGGED INTO AN ELECTRIC
OUTLET?
Make sure that the incubator is securely plugged in.
Does the output voltage at the electric outlet meet the power
requirement of the incubator?
CHECK THE CIRCUIT BREAKER INSTALLED IN THE ROOM OR
THE BUILDING FOR THE RATED VOLTAGE.
IS THE CIRCUIT BREAKER OF THE INCUBATOR THROWN?
Push the breaker back in and try switching on the incubator again.
Is the power switch thrown to the ON side?
PUSH THE SWITCH TO THE ON SIDE.
Power is on, but the circuit
breaker trips.
Is there any liquid inside the electric section of the incubator?
Should water or any other liquid poured or splashed onto the
incubator, unplug the unit immediately and contact us or our sales
representative. If left unserviced it could cause a fire.
IS THE INCUBATOR SITUATED IN A DUSTY PLACE?
Dust might have accumulated on the power cable, causing
short-circuit. Unplug the incubator immediately and contact us or our
sales representative. If left unserviced it could cause a fire.
The circuit breaker of the
building trips.
Is the incubator sharing the same power supply with many of other
instruments?
Plug the incubator into the isolated power supply. Continuing to use
the product in such condition could lead to a fire.
Is the incubator sharing the same power supply with another
instrument that requires large amount of power?
Plug the incubator into the isolated power supply.
Temperature inside the
incubation chamber is
unstable.
IS THE LID OF THE INCUBATION CHAMBER CLOSED
COMPLETELY?
Make sure that the lid is closed securely.
Temperature inside the
incubation chamber does not
reach the target temperature.
IS THE LID OF THE INCUBATION CHAMBER CLOSED
COMPLETELY?
Make sure that the lid is closed securely.
IS THE TARGET TEMPERATURE TOO LOW?
The target temperature must be higher than the room temperature by
C or more.