26. Precaution for set up
【The place to install】
Incubator chamber is maintained at stable temperature and gas concentration to make near-biological
body environment. The items below are the key points of the place to install.
A) The place to install must be firm and able to maintain horizontal
B) The room to install must be as clean as possible
(We recommend Clean Room to install and prior check of cleanliness.)
C) Do not install Drawer Type Incubator AD-3100 at places where
a. The wind blows against the incubator directly or indirectly such as around door, air conditioning
equipment and/or clean bench.
b. Are exposed to direct sunshine and/or wet location
c. Room temperature goes up and down precipitously
D) Connect the incubator only to a suitable power source specifically assigned for the incubator. Using
the correct power source specifically assigned for the incubator is extremely required.