EasyManua.ls Logo

ASTEC AD-3100 - O2 Calibration Procedures

ASTEC AD-3100
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AD-3100 Drawer Type Incubator Operation Manual Rev. 4
ASTEC CO., LTD.
24
O
2
ZERO Adjustment A-6
When measured O
2
concentration is different from the O
2
display concentration and it needs to be
adjusted, calculate the difference between these two values (A) and change the current value in A-6
by the same amount using <UP> and <DOWN> cursors. After changing the A-6 value, go back to
the normal display and confirm that O
2
display shows the same value as that of the measured O
2
concentration.
(A) = (Measured O
2
Concentration) (O
2
Display Concentration)
O
2
SPAN Manual Calibration A-7
This function allows the incubator to carry out manual ZERO Calibration for O
2
concentration. As a
default, A-7 shows OFF on the display. Change this to ON with <UP> and <DOWN> cursors and
wait for 10 second until the incubator starts manual calibration.
Make sure to use a calibrated thermistor when temperature in chamber is re-calibrated.
Do not change the value on A-2 and b1~Gb as these setting are only used by the
manufacturer.

Table of Contents

Related product manuals