EasyManua.ls Logo

Barco MDSC-8232 - Cleaning and Disinfection

Barco MDSC-8232
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
K5902094 (451920611693) /02 MDSC-8232
53
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Recommended separation distance
The MDSC-8232 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer of the user of the MDSC-8232 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the MDSC-8232 as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter
4
W
Separation distance according to frequency of transmitter
150kHz to 80MHz
d=1.2P
80MHz to 800MHz
d=1.2P
800MHz to 2.5GHz
d=2.3P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection form structures, object and people.
6.6 Cleaning and disinfection
Instructions
Be sure to unplug the power cord from the mains when cleaning your LCD monitor.
Take care not to scratch the front surface with any hard or abrasive material.
Dust, finger marks, grease etc. can be removed with a soft damp cloth (a small amount of mild detergent
can be used on the damp cloth).
Wipe off water drop immediately.
Possible cleaning solutions
250 ppm chlorine solution
NaCI solution 0.9% Sodim chloride 00-236
Bacillol AF
1.6 percent aqueous ammonia
Cidex® (2.4 percent glutaraldehyde solution)
Sodium hypochlorite (bleach) 10 percent
Green soap (USP)
Like Cleansafe® optical cleaning liquid
Isopropanol
Haemosol solution (1% in 1 liter water)
Chlorehexidine 0,5% in 70% Ethanol
Important information
4: For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter. Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.

Table of Contents