5. Charging is indicated by a permanently lit red LED located between the two buttons.
NOTE
Both the permanently lit red LED in the handheld and the flashing red LED in the
charger merely indicate the battery charging process and do not represent a
warning!
With empty NiMH batteries, the handheld can be charged for a short time using
a charging time of at least 15 minutes.
If the handheld is left in storage for longer periods with batteries inserted, the
batteries may discharge. If a prolonged storage without use is foreseeable, it is
recommended to remove the batteries from the handheld. After a storage period
of the handheld with inserted rechargeable batteries of more than 3 months, it
is recommended to fully recharge the rechargeable batteries before reuse.
If the batteries have discharged after a full charge and a too low battery voltage
is displayed, the batteries have become unusable. In that case, replace the bat-
teries.
During the charging process, elevated temperatures may occur in the handle
area of the handheld and in the charging station, which, however, decay imme-
diately after the charging process has been completed and do not pose any risk
to the user or patient.